Compact system for the production of contamination-free water from mains water. The device is connected directly to the mains water supply.
Simultaneous provision of ultrapure and pure water qualities:
Easy to use, space-saving design and low follow-up costs. The pre-treatment cartridge is separate from the ultrapure water cartridge. Only the used cartridge need be replaced, when necessary.
Applications:
Analytical chemistry:
HPLC, GC, GC-MS, ICP-MS, AA, TOC measurements, IC
Molecular biology and microbiology:
Cell/tissue culture, PCR, DNA sequencing, electrophoresis
Routine tasks in the laboratory
Preparation and dilution of buffers and reagents, rinsing of laboratory glass instruments, supply of autoclaves, analysers and other laboratory instruments
Drinking water requirements: | |
Drinking water to DIN 2000 | Manganese and iron content: <0.05 mg/l |
Chlorine concentration: <0.1 mg/l | |
Colloid index: <3 | |
Feed water temperature: 2-35 °C | |
Feed water pressure: 1-6 bar |
With UV lamp and UF module. The UF ultra-filtration module guarantees almost complete elimination of macro-molecular contamination such as micro-organisms, pyrogens (bacterial endotoxins), nucleases and particles; the combination of UV oxidation (254 and 185 nm) and ultra-filtration produces analytically pure water with a specific resistance of up to 18.2 MΩ cm, TOC values of <5 ppb, pyrogene levels of <0.001 EU/ml and RNase, DNase and DNA concentrations below the detection limit.Typical applications: molecular biology, PCR, DNA, monoclonal antibodies, cell culture media.
Type | Smart2Pure UV/UF 3 l/h |
Conductivity of high-purity water, type I | 0.055 µS/cm |
Conductivity of pure water, type II | 0,067-0,1 µS/cm |
Resistance of high-purity water, type I | 18.2 MΩ*cm |
Resistance of pure water, type II | 10-15 MΩ*cm |
Withdrawal capacity of high-purity water, type I | 0.6 l/min |
Withdrawal capacity of pure water, type II | 3 l/h |
Bacteria content of high-purity water, type I | <1 KbE/ml |
Endotoxin content of high-purity water, type I | 0.001 EU/ml |
TOC value of high-purity water, type I | 1-5 ppb |
Particle content of high-purity water, type I | <1 particle per ml (particle size <0.2 μm) |
Width | 305 mm |
Depth | 400 mm |
Height | 545 mm |
Weight | 22 kg |
Connection | Feedwater: hose (length 1.5 m) with Ø outside 8 mm and 3/4″ Waste water: 1/4″ or Ø outside 8 mm, hose (length 2 m), pressure-free discharge |
Mains connection | 100-240 V, 50/60 Hz |