For dry, clean and fast heating and mixing. For reaction vials, tubes and microtiter plates.
Programmable control:Programmable with 30 programming steps for up to 9 different programs. Program parameters: Nominal temperature, process time (max. 99 h/continuous), expansion time (time in which the nominal temperature is to be reached), shaking frequency, interval shaking, pause, etc.
Double heating thermoshaker with linear shaking motion creates a jerky motion for better mixing of small sample volumes in the ?l range. Suitable for two insertion blocks. Same position when stopped.
Type | MHL 23 |
Version | Heating |
Shaking motion | linear |
Max. exchange blocks | 2 |
Operating temperature | From 3 °C above room temperature to +130 °C |
Temperature adjustment range | 0 - 140 °C |
Accuracy | ±0,1 K |
Resolution | 0,1 °C |
Average heating rate | approx. 4 °C/min from +41 °Capprox. 5 °C/min up to +40 °C |
Shaking frequency | 200-1300 rpm |
Shaking amplitude (stroke) | 3 mm |
W x D x H | 220 x 345 x 160 mm |
Mains connection | 110/240 V, 50/60 Hz |